DC MICROAMPERES ANALOG METERS
Directions: For analog panel meters on Linaire test panels use a 3 1/2 digit
DVM with 1% accuracy or better to check the calibration of panel meters. Before
testing these meters the meter resistance must be checked. Set the meter load
switch on the Linaire panel to position 1 and check for 1000 ohms +/- 10 ohms
across the meter. (make sure that your DVM output is no more than 100millivolts
across a 1000 ohm resister first, anything more can damage the meter). If meter
resistance is out of limits most meters can be adjusted through a hole in the
back of meter case. Adjust the resistance as near to 1000 ohm as possible. If
resistance cannot be set properly the meter must be returned for repair or
replacement for a fee. A 0 - 500 uA current source into 1000 ohm load will be
needed for testing these meters. A dry cell battery, range resistors and a
potentiometer for fine current adjustment can be used as a satisfactory source.
Apply + voltage to meter + terminal for 0 - 500uA meter and 100-0-100uA meter
for right hand deflection. For left hand deflection reverse the supply leads.
Start with current source set to 0 uA and increase to meters max range.
Item: The microampere meter if found out of calibration can be returned to
Linaire for repair and recalibration or replacement for a fee.
Notes: : Adjust the meter pointer zero position carefully because this
can cause an error across the entire range of the meter. Meter should be tested
for smooth operation over the entire range of motion watching for places where
the pointer may stick, this is the most common problem to occur with these
meters as they age.
|